联系我们
  • 联系人 : 周先生
  • 联系电话 : 0512-68050030
  • 传真 : 0512-68050020
  • 移动电话 : 18901546660
  • 地址 : 苏州高新区狮山路75号
  • Email : 610421391@qq.com
  • 邮编 : 215011
  • 公司网址 : http://www.94817.com/
  • QQ : 610421391
产品详情
  • 产品名称:美国泛美Panametrics 35/35DL/35HP 超声波测厚仪

  • 产品型号:
  • 产品厂商:美国泛美panametrics
  • 产品价格:0
  • 折扣价格:0
  • 产品文档:
你添加了1件商品 查看购物车
简单介绍:
美国泛美Panametrics 35/35DL/35HP 全系列超声波测厚仪
详情介绍:

美国泛美Panametrics

The Panametrics-NDT™ Series 35 ultrasonic precision thickness gages provide easy-to-use and cost-effective solutions in applications where the opposite side of the test material is difficult or impossible to reach. These rugged, pocket-sized gages make stable, repeatable thickness measurements on most materials of varied shapes and sizes. Accurate thickness measurements are displayed in large numerals on the backlit LCD or can be viewed along with the live waveform in the optional A-scan Mode.

35 Series Features

• Velocity and Reduction Rate measurements are standard on all models
• Wide thickness range from 0.0030" to 25.0" (0.08 mm to 635.0 mm) depending on instrument and material
• Uses contact, delay line, and immersion transducers
• Application Auto-Recall with default and custom setups
• Hand-held; weighs only 8.5 oz. (0.24 kg)
• Min/Max Mode
• Hi-Low alarm
• English and metric display (inches/mm)
• Multi-language user interface
• Long battery life

Models 35 and 35DL

Models 35 and 35DLare used in the majority of applications
The 35 and 35DL can use transducers ranging from 2.25 to 30 MHz, which means that these versatile gages can solve the majority of thickness gaging applications, from very thin to very thick. In general, transducers with higher frequencies and smaller diameters allow measurements of thinner or curved parts and enhance the accuracy of the measurement.

Applications
• most materials, from thin to thick
• plastic bottles, tubes, pipes, sheets as thin as 0.08 mm (or 0.003 inch)
• metal containers, steel coils, machined parts as thin as 0.10 mm (or 0.004 inch)
• cylinder bores, turbine blades
• glass bulbs, bottles
• thin fiberglass, rubber, ceramics, and composite materials
• curved areas or containers with small radii
• resolution up to 0.001 mm or 0.0001"

Models 35HP and 35DL-HP

With a very low ultrasonic frequency bandwidth and a special pulser-receiver, the HP gages are specifically designed to optimize ultrasound penetration when measuring thick, highly sound attenuating or sound scattering materials. Typically these materials cannot be measured with most other ultrasonic thickness gages.

Applications
• most thick or sound-attenuating materials
• thick cast metal parts
• thick rubber tires, belts
• fiberglass boat hulls, storage tanks
• composite panels
• resolution of 0.01 mm or 0.001"
of 0.01 mm or 0.001"

The 35HP gages are excellent tools to measure fiberglass or composite parts, from aerospace structures to boat hulls and storage tanks that require thickness control.

Why HP gages
For more than three decades, Panametrics-NDT thickness gage engineers have developed the HP (High Penetration) series specifically for very sound-attenuating or thick materials. Not only can these gages use transducers as low as 0.5 MHz but their electronics (pulser-receiver) are highly optimized to process signals at these low frequencies. The result is that the Panametrics-NDT gages have earned a reputation for superior performance in measuring thick rubber, fiberglass, composites, and other tough materials.

Other 35 Series Features

With reduction rate you can measure critical thickness metal thinning caused by bending.


Reduction Rate Measurements
Differential Mode and Reduction Rate Mode are standard features on all models. Differential Mode shows the thickness variation from a pre-set thickness value. Reduction Rate calculates and displays the percent of thickness reduction after a material thinning process. A typical application is automotive sheet steel that is bent and formed to make car body panels.

Velocity measurement mode with direct Velocity readout.

 

Material Sound Velocity Measurements
All 35 models have the capability to make material sound velocity measurements. This standard feature is useful in applications where the speed of the sound within the material can be correlated to other properties. Typical applications include cast metals to monitor the degree of nodularity, and composites/fiberglass to monitor variations in density. Olympus NDT offers a digital caliper for automatic transfer of measurement thickness.

Application Auto-Recall Simplifies Gaging
Application Auto-Recall simplifies making thickness measurements. Select any of the stored transducers and the 35 gages recall all relevant internal transducer parameters.
Stored Standard Setups
Standard setups include most commonly used transducers.
Stored Custom Setups
Just in case your special application problem cannot be solved with a Standard Setup, these gages can create, store, and recall as many as 20 Custom Setups (10 with the 35 and 35HP).

Micrometers won't work in this classic application. Instead, the Model 35 with a M208 transducer makes a thickness measurement up to a calibrated accuracy of 0.001 mm (or 0.0001 inch) without breaking the glass .

 

Thickness and waveform of a very thin (0.004 inch or 0.10 mm) sheet of steel.

 

35PCSCOPE

Optional 35PCSCOPE interface software permits the instant display of live ultrasonic waveforms and thickness readings directly on your computer screen. This is helpful when setup parameters require closer examination.

Now You Can See The Live Waveform And Make Adjustments

Operator can view thickness and waveform with the optional A-scan mode

An optional Live Waveform Mode on all models permits the user to view the ultrasound waveform (or A-scan) directly on the gage's display to make sure the thickness reading is correct. The Adjust Mode feature facilitates transducer setup adjustments to maximize measurement performance for challenging applications.
• High pixel density LCD renders sharp trace of live waveform
• Permits real-time verification of thickness readings
• Helps solve many difficult applications
• A Range, Delay, or Zoom allow enlarging and centering to any section of the waveform
• Permits adjustments of transducer setup parameters
• Stores waveforms in data logger and transfers to PC

Why ultrasonics?
Ultrasonic measurements are accurate, reliable, and repeatable. Instant digital readings can be achieved by transmitting sound into just one side of a material, making it unnecessary to cut or destroy parts where the opposite side is difficult to reach and where micrometers or other inspection tools cannot do the job.

35, 35DL, 35HP and 35DL-HP Specifications*

Measurements

Mode 1:

Time interval between excitation pulse and first back wall echo, using contact transducers.

Mode 2:

Time interval between the first interface echo after the excitation pulse and the first backwall echo, using delay line or immersion transducers.

Mode 3:

Time interval between successive back wall echoes following the first interface echo after the excitation pulse, using delay line and immersion transducers.


Measurement Types: Thickness, Velocity, or Time of Flight
ThicknessMeasurement Range*:
0.003 - 25 inches (0.08 - 635 mm)
*Thickness range depends on model, material, transducer, surface condition, and setup selected.

MaterialVelocity Range:
0.02000 - 0.7362 inch/µs
0.5080 - 18.699 mm/µs)

Resolution (keypad selectable):

LOW:

0.01"

(0.1 mm)

STANDARD:

0.001"

(0.01 mm)

HIGH:

0.0001"

(0.001 mm)
(35 & 35DL)

Time of Flight Measurement Range:
0.0 - 109.5 µs

Time of Flight Resolution:
Fixed at 000.01 µs

Measurement Update Rate:
4, 8, 16, or Max Hz (16-20 Hz depending on application and measurement mode)

TransducerFrequency Range:
2.25 - 30 MHz (35 & 35DL)
0.5 - 5.0 MHz (35HP & 35DL-HP)

Min/Max Mode: Displays current thickness, minimum or maximum thickness depending on setting.

Display Hold/Blank: Displays blanks after last reading or holds reading.
Alarm Mode: Programmable Hi-Low alarm set points with audible and visual indicators

Differential Mode: Displays thickness difference between actual measurement and reference value.

Reduction Rate Mode: Displays thickness and percent difference between actual measurement and reference value.

Application Auto-Recall: Automatically adjusts internal parameters for various default and custom transducer setups.

 

# of Default

# of Custom

35

21

10

35DL

21

20

35HP

17

10

35DL-HP

17

20

Other Standard Features: Calibration Lock Mode, Internal Diagnostic Test Mode, Metric/English Units, On-board Statistics (35DL and 35DL-HP only)

POWER SUPPLY
Battery: 3 AA alkaline or NiMH batteries
Operating Time: 150 hours typical battery life; 30 hours continuous with backlight on
Low Battery Indicator: Continuously indicates battery status.
BatterySaver: Auto Power Off/Continuous On
AC Power Supply: Optional

GENERAL
Environmental IP-65 compliant:
Splash-proof, impact-resistant case. Sealed, color-coded keypad with tactile and audible feedback.
Operating Temperature Range:
-10°C to +50°C (+14°F to 122°F)
Size: 3.31" W x 6.0" L x 1.50" H
(84 x 152 x 84 mm)
Weight: 8.5 oz. (0.24 kg)
Custom Language Utility: Allows translation of display text in many languages.

35DL & 35DL-HP
INTERNAL DATA LOGGER
Data Logger: The 35DL & 35DL-HP identify, store, recall, clear, and transmit thickness readings and transducer setup information via the USB port or optional RS-232.
Max. # of Stored Values: Over 8,000 or 350 waveforms with thickness readings (with Waveform Option)
Stored Data Documentation: Each saved thickness reading is fully documented with measurement status flags and a setup number that identifies parameters such as velocity, transducer, etc.
File Name Length: 8 alphanumeric characters
Identification Codes: 10 character alphanumeric Identification Code system identifies or locates stored data
4 File Templates: Incremental, Sequential and 2D Grid. Manual files from PC.

STANDARD INCLUSIONS
Wrist Strap, 3 AA Batteries, Test Block, Transducer Cable, Couplant, Carrying Case, Instruction Manual, Two-Year Limited Warranty

OPTIONAL ACCESSORIES

35/WF

Live Waveform

Gageview

Interface Program

35DL/RS232

RS-232 output

35DLHP/RS232

RS-232 output

35PCScope

PCScope Program

USB/ADP-115

AC-115 Power Supply

USB/ADP-230

AC-230 Power Supply

MG2/RPC

Protective Rubber Boot

35/DLRETRO

Upgrade 35 to 35DL

35HP/DLRETRO

Upgrade 35HP to 35DL-HP

*All specifications are subject to change without notice.

Some 35 Series Applications

 

 

Wall Thickness Measurements for Metal Cans, Containers and Enclosures

Thickness Inspection of Metal Coils

Measuring Wall Thickness of Gun Drilled Metal

Ultrasonic Thickness Measurements of Hot Metals

Wall Thickness Measurements of Metal Pipes and Tubes

On-line Thickness Gauging of Extruded Plastic

High Temperature Ultrasonic Testing

Ultrasonic Couplants

Measurements of Silicone Coatings on Porcelain High Voltage Insulators

Base Curve and Thickness Measurement of Soft Contact Lenses

Ultrasonic Measurements of Liquids

Thickness Measurement of Glass

Thickness Testing of Reinforced Rubber Conveyor Belts

Measuring Thickness of Rubber Tires

Wall Thickness Measurements of Rubber Tires and Bladders

Thickness Measurements on Rubber Sheets and Coatings

Thickness Measurements on Rubber Hoses & Tubing

Measurement of Fiberglass Gelcoat

Thickness Gaging of Fiberglass Boat Hulls

Aerospace Composites

Thickness Measurements of Fiberglass Plates, Pipes and Storage Tanks

Measuring Nodularity in Cast Iron

Measuring Thickness of Wax Molds for Precision Castings

Wall Thickness Measurement of Cast Engine Blocks and Cylinder Bores

Iron, Steel, Aluminum, and Non-Ferrous Castings

Ultrasonic Testing in the Foundry Industry

Wall Thickness Gaging in the Blow Molding Industry

Elastic Modulus Measurement

 

标题:
内容:
联系人:
联系电话:
Email:
公司名称:
联系地址:
 
 
注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
2.如有必要,请您留下您的详细联系方式!

苏公网安备 32050502000402号